Specification
HF4500
Product introduction
Summary

Device Overview

The new HF4500 microplate reader is developed based on extensive experiences in practical application. The instrument incorporates the latest technologies and strengths of similar domestic and international products. The instrument performs both single-wavelength and dual-wavelength assays, as well as qualitative and quantitative enzyme-linked immunoassays.

Equipment classification: Temporary pressured equipment; Category II; Pollution level: Class 2

Technical specifications

Light source: halogen, automatic calibration

Measurement channels: 4 vertical, optical channels; automatic centering

Wavelength range: 405 nm (optional), 450 nm, 492 nm, 630 nm; error of central wave:± 2 nm

Microplate type: 2X12 4X12 8X12

Measurement range: 0.000 to 4.000 OD

Resolution: 0.001 OD

Linear range: 0.000 to 3.000

Repeatability: 0.001

Accuracy: ±0.010

Display and input: 7-inch color LCD display; touch screen input?

Power input: wide voltage range of AC 220 V to 240 V, 50 to 60 Hz?

Product weight: Main unit has a net weight of 5.7 kg?

Dimensions: 390 mm (L) x 270 mm (W) x 155 mm (H)


Precautions
1) There should be a reliable power supply within the range of AC 220 V ± 10%, the frequency should be within 50 Hz ± 2%. Do not connect to 380 V voltages, and do not turn off the power during use. 2) Keep away from light, and store in a clean indoor environment away from strong electromagnetic fields. 3)?Maintain the required humidity (30% -80%), and maintain a relatively constant temperature in the room (20-25℃ is most suitable). 4)?Please use the standard 96-well ELISA plate for reading. 5)?Do not leave the micro-well plate and cuvette in the instrument slot for too long time after the test. Remove them from the instrument promptly after the test to prevent the solution from evaporating and corroding or damaging the internal optical path of the instrument. If a corrosive or volatile solution is tested, measure with a cover attached. 6)?Should the instrument behave abnormally or does not work properly, contact the manufacturer immediately. Do not disassemble or adjust the instrument. 7)??The instrument can be cleaned by wiping it with 2% alkaline or neutral glutaraldehyde solution or 0.5% chlorhexidine acetate-ethanol solution. 8)?In case of serious contamination, the instrument can be disinfected by the ethylene oxide disinfection method. 9)?Please follow the printer instructions to select acceptable printing paper.